Recall Z-0281-2023 ยท OWN Confocal Optical Imaging

Visionsense, Ltd.

VS3 Iridium - Visionsense Infrared (IR) Fluorescence Miniature Microscope (MMS) - 785nm Product Number: 175-0012

DateNov 23, 2022 (initiated Oct 17, 2022)
StatusOpen, Classified
ReasonXXX
Root causeUnder Investigation by firm
ActionMedtronic issued Urgent Medical Device Correction Letter on 17-OCTOBER-2022 to US consignees via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and action to take: 1.Prior to use, visually check the Beam-Combiner and MMS optical fiber cable for signs of being pinched, kinkedor damaged. Avoid folding the cables and handle with care to prevent damage. 2.After the white balance has been performed as per IFU guidance, test the EleVision IR system at the beginningof the surgical procedure as
Quantity34 units: 15 units OUS; 19 units US
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150018VS3-IR-MMS System
Visionsense, Ltd., Jun 17, 2015