Recall Z-0350-2023 ยท JPK Mixture, Hematology Quality Control

Primus Corporation dba Trinity Biotech

Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis

DateDec 1, 2022 (initiated Oct 5, 2021)
StatusTerminated
ReasonThe Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.
Root causeNonconforming Material/Component
ActionThe firm sent an Urgent Filed Safety Notice their consignees on 10/05/2021 by mail. The notice explained the issue, risk and requested the destruction of of the affected devices.
Quantity182 kits
DistributionTurkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.