ClearanceBrief
Recall Z-0225-2023 ยท FBK Endoscopic Injection Needle, Gastroenterology-Urology

Boston Scientific Corporation

ORISE Gel Syringe Submucosal Lifting Agent Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device UPN: (1)Twin Pack Box 1 M00519210; (2)Twin Pack Box 10 M00519211; (3)Si

DateNov 11, 2022 (initiated Oct 10, 2022)
StatusOpen, Classified
Reason***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention
Root causeUnder Investigation by firm
Action***UPDATE 12/13/22** Boston Scientific issued Urgent Medical Device Product Removal Immediate Action Required Orise" Gel Submucosal Lifting Agent via Federal Express Priority Overnight mail on December 14, 2022, (as a Follow-Up To The Previous Product Advisory Dated October 10, 2022). Letter states reason for recall (Product Removal), health risk and action to take: Use and distribution of any remaining unused ORISE Gel Submucosal Lifting Agent products affected by this removal should cease immediately. For patients already treated with ORISE Gel, should the user identify submucosal distor
Quantity402,968 Total
DistributionWorldwide Distribution: US (Nationwide) and (Foreign) countries of: Algeria Argentina Argentina Australia Austria Bahrain Belgium Brazil Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt Finland France Germany Great Britain Greece Hong Kong Hungary Iceland Iraq Ireland Isr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171454Interject Injection Therapy Needle Catheter
Boston Scientific Corporation, Jul 7, 2017