Recall Z-0198-2023 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic
Canon Medical System, USA, INC.
Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450
| Date | Nov 8, 2022 (initiated Sep 29, 2022) |
| Status | Open, Classified |
| Reason | After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay. |
| Root cause | Software Design Change |
| Action | On 9/29/22, correction notices were mailed to customers who were asked to do the following until the software corrective action had been implemented: Terminate Stress Echo by pressing "X" or "Quit" on the touch panel when completing the exam. If the exam is completed without terminating Stress Echo, and the confirmation message shown in section 2 is displayed when saving the first image in the next Stress Echo, terminate Stress Echo by pressing "X" or "Quit", and then start Stress Echo again. However, if the confirmation message is not displayed when saving the first image in Stress Echo, the |
| Quantity | 65 |
| Distribution | US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL |