Recall Z-0198-2023 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Canon Medical System, USA, INC.

Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450

DateNov 8, 2022 (initiated Sep 29, 2022)
StatusOpen, Classified
ReasonAfter completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.
Root causeSoftware Design Change
ActionOn 9/29/22, correction notices were mailed to customers who were asked to do the following until the software corrective action had been implemented: Terminate Stress Echo by pressing "X" or "Quit" on the touch panel when completing the exam. If the exam is completed without terminating Stress Echo, and the confirmation message shown in section 2 is displayed when saving the first image in the next Stress Echo, terminate Stress Echo by pressing "X" or "Quit", and then start Stress Echo again. However, if the confirmation message is not displayed when saving the first image in Stress Echo, the
Quantity65
DistributionUS: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212333Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V6.5
Canon Medical Systems Corporation, Jan 24, 2022
K212960Aplio a550, Aplio a450, and Aplio a, Diagnostic Ultrasound System, Software V6.5
Canon Medical Systems Corporation, Mar 22, 2022