Recall Z-0221-2023 ยท LPC Device, Angioplasty, Laser, Coronary

Spectranetics Corporation

Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular system, and for the removal of problematic pacemaker and defibrillator cardiac leads.

DateNov 11, 2022 (initiated Oct 14, 2022)
StatusOpen, Classified
ReasonLaser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation of treatment delay and/or treatment beyond initial scope and/or unable to treat patient.
Root causeDevice Design
ActionOn 10/14/22, correction notices were mailed to customers who were asked to take the following actions: Firm recommends continued use of affected devices and to follow the operator's manual. As a temporary mitigation, until software becomes available tentatively in Q2 2023, the user can perform steps when visually alerted on the system display as follows: Error 106 Firm's internal testing confirms that the following steps may temporarily resolve the issue, if encountered: 1. System restart 2. Recalibration of the catheter in use or calibration of a new catheter Error 108 Firm's internal tes
Quantity113
DistributionUS: LA, MO, TN, CA, NC, PA, HI, AZ, CO, OK, FL, TX, MN, MI, VA, CT, NY, MD, IL, NJ, MA, KS, ME, AR, IA, IN, GA, ID, OR. OUS: DE, GB, ES, SE, IT, RS, GR, IN, FR, IL, DK, SA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.