Recall Z-0184-2023 ยท FPO Stretcher, Wheeled
Ferno-Washington Inc
FERNO POWER X1 AMBULANCE COT- Item #0015807/UDI-DI 00190790002183 with 36V Li-Ion Batteries Model Number #0822484/0822483 Note: Transcend Stair Chair Powertraxx (Item #0731403/UDI-DI 00190790004064) not included in this action).
| Date | Nov 4, 2022 (initiated Oct 7, 2022) |
| Status | Open, Classified |
| Reason | Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing directly into battery contacts or abuse, may result in a fire and/or injury |
| Root cause | Component design/selection |
| Action | Ferno issued Urgent Product Correction on 10/7/22 via Certified Mail. Letter states reason for recall, health risk and action to take: Ferno will replace affected batteries with new batteries. Immediate Action. 1. Immediately check the serial number of your Ferno POWER X1 Ambulance Cot (see list within this notice) as well as the serial number on the bottom of any ENDURACharge 36V Li-Ion battery utilized with the cot. See How to Identify ENDURACharge Battery Date Code . 2. Check the serial number on the bottom of your ENDURACharge 36V Li-Ion battery to see if the battery is in the identified |
| Quantity | 1,922 units |
| Distribution | Nationwide Foreign: BRAZIL BRUNEI CHILE CHINA PERU POLAND PUERTO RICO TURKS&CACOS |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.