Recall Z-0229-2023 ยท CAW Generator, Oxygen, Portable

DeVilbiss Healthcare LLC

125 series iGO2 Portable Oxygen Concentrator-indicated for the administration of supplemental oxygen. Models: 125D, 125D-XB

DateNov 11, 2022 (initiated Sep 19, 2022)
StatusOpen, Classified
ReasonAn unapproved adhesive used to manufacture the accumulator tank sub assembly (a part of the patient gas pathway)
Root causeUnder Investigation by firm
ActionDeVilbiss notified Distributors/End Users by telephone on 9/19/22. Providers of portable oxygen concentrators, identified as recipient of affected units, were contacted asking them to return any finished good units in their possession with affected serial numbers to DeVilbiss for evaluation. A replacement unit will be provided once unit possession and shipping confirmation is received. If any unit has been supplied by the provider to an end user, the provider is being requested to contact the end user to request them to stop using the unit and to replace the unit immediately. If you have an
Quantity48 units
DistributionUS Distribution in states of: CA, FL, OH, MI, MN, NY, PA, and TX; and (Foreign) country of: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172648Drive DeVilbiss iGo 2 Portable Oxygen Concentrator