Recall Z-0181-2023 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Fresenius Medical Care Holdings, Inc.

NaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case. For dialysis.

DateNov 4, 2022 (initiated Oct 4, 2022)
StatusOpen, Classified
ReasonThe "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.
Root causeError in labeling
ActionOn October 4, 2022, the firm notified affected customers via Important Medical Device Information letters. Customers were reassured that the product was manufactured according to specification and remains potassium-free. Customers may continue to use the product. Please contact FMCRTG Customer Service at 1-800-323-5188 if you have further questions.
Quantity17700
DistributionDomestic distribution to AL AZ CA DC FL GA HI IL IN MD MI MO MS NE NJ NY OH PA SD TX VA No foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070177FRESENIUS NATURALYTE LIQUID ACID CONCENTRATE, 9000, 6000 AND 4000 SERIES
Fresenius Medical Care North America, Mar 29, 2007