ClearanceBrief
Recall Z-0117-2023 ยท NKM Mitral Valve Repair Devices

Abbott Vascular

Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT, MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT, TriClip G4 NT Delivery System CLIN/TCDS0303-NT, TriClip G4 NTW Delivery System CLIN/

DateOct 20, 2022 (initiated Sep 8, 2022)
StatusOpen, Classified
ReasonAn increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
Root causeProcess control
ActionOn 9/8/22, correction notices were distributed to customers who were asked to do the following: 1) Read and share the information in the notice with all device users. 2) continue to follow the steps of the IFU (eifu.abbottvascular.com) that is also summarized in the correction notice. It details the most relevant steps to aid in optimal function of the Clip locking mechanism, thereby reducing the probability of EFAA failure and COWL to occur. 3) Complete and return the Acknowledgement Form. 4) If you have transferred affected products to other centers, inform those centers of this notice. Th
Quantity767
DistributionWorldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.