Recall Z-0267-2023 ยท CBK Ventilator, Continuous, Facility Use

Hamilton Medical AG

Hamilton-C6

DateNov 18, 2022 (initiated Sep 22, 2022)
StatusOpen, Classified
ReasonDue to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.
Root causeUnder Investigation by firm
ActionOn 10/19/2022, Hamilton Medical, Inc. sent an "URGENT - Medical Device Correction" Letter via email to customers informing them the firm had received a complaint report associated with the backlight of the HAMILTON C-6 display failing. The failure is attributed to a reset of the HMI controller reset with causes the display to fail for 2-3 seconds. Hamilton Medical is notifying customer that this failure does not affect or interrupt the device's ventilation function. Customers are instructed to: -Sign the enclosed sheet confirming that they have received and understood this Medical Device Co
Quantity8,229 total systems worldwide (U.S. 443 systems)
DistributionU.S. Nationwide distribution including in the states of CA, FL, GA, ID, IN, KS, KY, MA, MI, MO, NV, NM, NY, NC, OH, PA, Puerto Rico, SC, TX, VA, WA, WV, WI, and WY.