Recall Z-0230-2023 ยท NQO Prosthesis, Spinous Process Spacer/Plate

Boston Scientific Neuromodulation Corporation

Superion Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm

DateNov 14, 2022 (initiated Oct 12, 2022)
StatusOpen, Classified
ReasonDue to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.
Root causeProcess change control
ActionOn 10/12/2022, the firm, Boston Scientific, sent an "Urgent: Medical Device Recall" letter via FedEx to customers. The letter described the product, problem and actions to be taken. The Customers are instructed to: -Review their facility's device inventory: Segregate and return any unused Superion IDS devices from the affected subset to Boston Scientific. Complete and return account-specific Reply Verification Tracking Form via Email: BSCFieldActionCenter@bsci.com or Fax to: Field Action Center at 1-763-415-7708. -Patient follow-up: Discuss this advisory with their patient(s) who are implant
Quantity175 devices
DistributionU.S Nationwide Distribution including states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MI, MN, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TX, UT, WI, WV, and WY
FDA recordOpen on fda.gov