Recall Z-0193-2023 ยท OZP Automated Insulin Dosing Device System, Single Hormonal Control

Medtronic MiniMed

Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742

DateNov 8, 2022 (initiated Sep 20, 2022)
StatusOpen, Classified
ReasonMedtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.
Root causeComponent change control
ActionOn 09/20/2022, correction notices were distributed to customers, healthcare professionals, distributors, and service providers. They were informed to take the following actions: 1. Turn off the "Remote Bolus" feature on your pump if it is turned on. Note that the "Remote Bolus" capability is on by default, so take this action even if this feature has never been used. See Appendix: "How to Turn Off Remote Bolus Settings" or by visiting our website at www.medtronicdiabetes.com/remote-bolus 2. Conduct any connection linking of devices in a non-public place. 3. Complete and return the customer co
Quantity399651
DistributionUS: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: Andorra, Australia, Austria, Barbados, Belgium, Bermuda, Bolivia, Botswana
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.