Recall Z-0201-2023 ยท GAM Suture, Absorbable, Synthetic, Polyglycolic Acid

Covidien, LP

Endo Stitch V-Loc PBT Non-Absorbable Reload, Models VLOCN004L, VLOCN006L, VLOCN008L, VLOCN204L, VLOCN206L, VLOCN208L, VLOCN304L, VLOCN306L, VLOCN308L

DateNov 11, 2022 (initiated Sep 28, 2022)
StatusOpen, Classified
ReasonMedtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.
Root causeProcess control
ActionBeginning 28-September-2022 communications to all affected consignees began. Consignees are asked to return any affected product they may have, forward this notification to those who need to be aware within the organization and to any location which the devices have been transferred. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity30658
DistributionDomestic distribution nationwide. Foreign distribution to Australia Austria Belgium Canada Canary Islands Chile Croatia Cyprus Denmark Egypt Finland France Germany Greece Hungary Ireland Israel Italy Jordan Kazakhstan Kuwait Lithuania Luxembourg Mexico Netherlands New Zealand Norway Poland Portugal