Recall Z-0078-2023 ยท HHK Curette, Suction, Endometrial (And Accessories)

Olympus Corporation of the Americas

Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.

DateOct 12, 2022 (initiated Sep 16, 2022)
StatusOpen, Classified
ReasonUpdate to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters which were not designed for use with the VC-10, which can lead to serious injuries as it does not fit properly with the VC-10.
Root causeLabeling design
ActionOn September 16, 2022, the recalling firm notified affected customers through "URGENT Medical Device Field Corrective Action" letters. Customers were informed that the following new warning will be added to the IFU: Do not use replacement parts or accessories that are not compatible with the VC-10 system. Use of noncompatible accessories could lead to reduced system functionality and/or patient injury such as perforation, bleeding, and may require additional intervention. Refer to Section 11 for all accessories compatible with the VC-10 system. Customers should notify their users about risks
Quantity4184 US; 191 OUS
DistributionDomestic distribution nationwide. Foreign distribution to Canada and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171440Berkeley VC-10 Vacuum Curettage System
Gyrus Acmi, Sep 29, 2017