Recall Z-0089-2023 ยท JHM Colorimetric Method, Lipoproteins

Siemens Healthcare Diagnostics, Inc.

Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A

DateOct 14, 2022 (initiated Aug 24, 2022)
StatusOpen, Classified
ReasonDimension Vista High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.
Root causeUnder Investigation by firm
ActionSiemens issued an Urgent Medical Device Correction (UMDC) mailed to US affected customers by FedEx beginning on 08/24/2022. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 08/24/2022. Letter states: Discontinue use of and discard the lot listed in Table 1. Review your inventory of this lot to determine your laboratory s replacement needs and to provide information to Siemens for reporting to the Authorities. Please review this letter with yo
Quantity1592 units
DistributionNationwide Foreign: Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081300DIMENSION VISTA HDLC FLEX REAGENT CARTRIDGE
Siemens Healthcare Diagnostics, Jul 2, 2008