Recall Z-0172-2023 ยท GKZ Counter, Differential Cell

Becton, Dickinson and Company, BD Biosciences

BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.

DateNov 3, 2022 (initiated Oct 3, 2022)
StatusOpen, Classified
ReasonDue to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
Root causeUnder Investigation by firm
ActionOn 10/03/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx tracked letters and emails informing customers, through internal investigations, the labels applied to the BD Trucount tubes are becoming detached due to an adhesive change made at a supplier. Customers are instructed to: 1. Immediately review their inventory for the specific catalog and lot numbers. 2. For any tubes where label detachment is observed, BD recommends that the clinician remove the label from the tube and, using a permanent marker, record the tube lot number as well as any other relevant informat
Quantity11,709 tubes (corrected 11/04/2022)
DistributionU.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090967BD MULTITEST 6-COLOR TBNK REAGENT, MODEL 644611
Bd Biosciences, Jul 31, 2009