Recall Z-0145-2023 ยท KPN Alarm, Conditioned Response Enuresis

Baxter Healthcare Corporation

Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for VersaCare Bed Rev. A-J, Product Code/Part Number P00697901

DateOct 31, 2022 (initiated Sep 30, 2022)
StatusOpen, Classified
ReasonRadio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Root causeDevice Design
ActionBaxter issued an Urgent Medical Device Correction on 09/30/2022. An updated letter was sent on 10/21/2022 to request that users temporarily stop using the WatchCare system. Hillrom, a Baxter company, has received reports of Radio Frequency (RF) interference with other devices near beds that are installed with WatchCare. Specifically, RF interference has been reported with telemetry devices, a bladder scanner, a fetal monitor/doppler, an infusion pump, and a personal use insulin pump. Other third-party medical devices may also be affected. While the system was developed to comply with the most-
Quantity129 units
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.