ClearanceBrief
Recall Z-0176-2023 ยท GDC Table, Operating-Room, Electrical

Baxter Healthcare Corporation

C-Flex Polar Head Positnr-Device Only, SSP-NEUROSURGICAL HEADREST, REF A-70701, #103001,

DateNov 3, 2022 (initiated Sep 20, 2022)
StatusTerminated
Reasonthe adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
Root causeOther
ActionThe firm issued an Urgent Medical Device Correction notice to their consignee on 09/20/2022 by email. The notice explained the problem, risk, and requested the following: "Actions to be taken by Customer: 1. Identify the impacted lot number listed on page 1 and stop any use of the product immediately. 2. Hillrom will contact you to perform a correction of your impacted device as soon as possible." The firm issued an Urgent Medical Device Recall notice to their consignees on 09/27/2022 by FedEx 2nd day delivery. The notice explained the problem, risk, and requested the following: "Actions t
Quantity2 units
DistributionMI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.