Recall Z-0086-2023 ยท HQC Unit, Phacofragmentation

Johnson & Johnson Surgical Vision, Inc.

Veritas Advanced Infusion Packs. Used in ophthalmic procedures.

DateOct 12, 2022 (initiated Sep 1, 2022)
StatusOpen, Classified
ReasonAs a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.
Root causeUnder Investigation by firm
ActionOn 09/01-02/2022, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via UPS 2-Day Mail to customers informing them that the firm is recalling specific lots of VERITAS Advanced Infusion Packs due to the potential for the irrigation luer to crack or break which could lead to reduced irrigation pressure during surgery, associated with an unstable anterior chamber. Customers are required to: 1.Identify if any of their inventory contains VERITAS Packs with a lot number listed on page 5. 2.Immediately discontinue using and remove from their inventory all affected VERITAS Packs. No other VERIT
Quantity53,624 units
DistributionWorldwide Distribution. US states of AL, AR, AZ, CA, CO, DE, FL, GA, HI, IL IN, KY LA, MN, MO, MS, NC, NJ, NY, OH, TN, TX, UT, VA, and WA; Australia, Azerbaijan, Belgium, Brazil, Canada, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, India, Iraq, Ireland, Italy, Japan, Jordan, Kuwa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203060VERITAS Phacoemulsification Console, VERITAS Advanced Fluidics Pack and Advanced Infusion Pack, VERITAS Swivel Handpiece
Johnson & Johnson Surgical Vision, Inc., Mar 12, 2021