ClearanceBrief
Recall Z-0085-2023 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corporation

Prismaflex HF20 Set (product code 109841). For use in providing continuous fluid management and renal replacement therapies.

DateOct 12, 2022 (initiated Sep 14, 2022)
StatusOpen, Classified
ReasonWithin the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Root causeError in labeling
ActionBaxter sent an Urgent Medical Device Correction notice to its direct consignees by USPS on 09/14/2022. The notice explained the Estonian translation issue and requested that the following actions be taken by consumers: 1. Use of the Prismaflex HF20 set should be restricted to patients with a body weight greater than 8kg (18lb). Use of the Prismaflex M60 and ST60 sets should be restricted to patients with a body weight greater than 11kg (24lb). Use of the Prismaflex M100, ST100, M150, ST150, HF1000 and HF1400 sets should be restricted to patients with a body weight greater than 30kg (66lb). 2.
Quantity29,764 devices
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042938GAMBRO PRISMAFLEX HF1000 AND HF1400 SETS
Gambro Renal Products, Jan 6, 2005