Recall Z-1786-2022 ยท IZL System, X-Ray, Mobile

Micro-X Ltd.

MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

DateSep 23, 2022 (initiated Aug 31, 2022)
StatusOpen, Classified
ReasonSoftware calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
Root causeProcess control
ActionOn or about 09/31/2022, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email/telephone to customers informing them that during an internal data analysis on 08/18/2022, Micro-X Ltd identified software calibration errors with the product equipped with a Dose Area Product (DAP) meter. This resulted in the product displaying incorrect DAP meter values. The DICOM image produced by the system will report a DAP measurement in microgray/square meter that is 8 to 10 times lower than the actual dose delivered. Customers are instructed to: 1. Inspect their stock immediately and place
Quantity42 units
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211423Rover
Micro-X Limited, May 21, 2021