Recall Z-1816-2022 · JGJ Photometric Method, Magnesium

Siemens Healthcare Diagnostics, Inc.

Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57

DateSep 26, 2022 (initiated Aug 17, 2022)
StatusOpen, Classified
ReasonDimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in the absence of an Abnormal Reaction flag. Based on customer data, imprecision leading to erroneous, unflagged patient results with a bias of -15 to -59% may occur
Root causeUnder Investigation by firm
ActionSiemens Healthineers issued to US Customers Urgent Medical Device Correction (VC-22-01.B.US) on 8/17/22 via Fed'X and OUS Customers sent Urgent Field Safety Notice (VC-22-01.B.OUS) to inform them to discontinue use and discard Dimension MG lots FA2350, FA2356, GA2363, and FA3019. Letter states reason for recall, health risk and action t take: Discontinue use and discard Dimension MG lots listed in Table 1. Review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please revi
Quantity26851 units
DistributionNationwided Foreign: Afghanistan¿¿¿¿ Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Canada Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K861700DIMENSION (TM) CLINICAL CHEMISTRY SYSTEM
E.I. Dupont DE Nemours & Co., Inc., Jun 11, 1986