Recall Z-1744-2022 ยท MUJ System, Planning, Radiation Therapy Treatment

Philips Medical Systems (Cleveland) Inc

Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.

DateSep 15, 2022 (initiated Aug 30, 2022)
StatusOpen, Classified
ReasonWhen computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.
Root causeDevice Design
ActionThe recalling firm issued letters dated August 2022 via certified mail on 8/30/2022. A black box at the beginning of the letter notates the document contains important information for the continued safe and proper use of the product. The information is to be reviewed with all members of the consignee's staff who need to be aware of the contents of the communication and to understand the implications of the communication. The letter informs the consignee what the problem is and under what circumstances it can occur and contains an embedded photograph of the Edit Prescription screen highlight
Quantity169 systems
DistributionDistribution was made to CA, FL, GA, IA, IL, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, OK, RI, SC, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Australia, Austria, Belgium, China, France, Germany, Italy, Japan, Jordan, Lebanon, Pola
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170086Pinnacle Radiation Therapy Planning System
Philips Medical Systems (Cleveland), Inc., Feb 9, 2017