ClearanceBrief
Recall Z-1851-2022 ยท FSO Microscope, Surgical, General & Plastic Surgery

Leica Microsystems, Inc.

Leica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950

DateSep 29, 2022 (initiated Aug 26, 2022)
StatusOpen, Classified
ReasonDuring internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
Root causeComponent change control
ActionThe firm sent a Field Safety Notice to its consignees on 08/26/2022 by mail. The notice described the problem and stated that A Leica Microsystems representative will be contacting the consignee's to arrange this no-cost field update of the instrument. The field update will be scheduled to occur at your convenience within the next 6 months. The notice provided the following workaround: "Advice on action to be taken by the user: " You can continue to use the system pending this field update. Based on our analysis the probability is very remote that an unacceptable patient risk will occur du
Quantity32 devices
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.