ClearanceBrief
Recall Z-1820-2022 ยท JZO System, Test, Thyroid Autoantibody

Siemens Healthcare Diagnostics, Inc.

IMMULITE 2000 and IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) Siemens Material Number (SMN): 10876421 US; 10713448 (OUS)

DateSep 27, 2022 (initiated Aug 11, 2022)
StatusOpen, Classified
ReasonAn average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid
Root causeUnder Investigation by firm
ActionSiemens Healthcare issued an "Urgent Medical Device Correction (UMDC IMC 22-09.A.US)" to US customers, and an "Urgent Field Safety Notice (UFSN IMC 22-09.A.OUS)" to Outside the US (OUS) customers. The UMDC and UFSN letters provide instructions to customers to follow: When this issue occurs, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid. Mitigations would include correlation of tests results with additional tests such as anti-TPO antibodies along with thyroid function tests such as total T3 and T
Quantity100 units US; 328 units OUS
DistributionWorldwide Distribution: US (nationwide) to states of: FL IN MA MI NH NJ NY OH TX UT and OUS (Foreign) countries of: Aus
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152061IMMULITE 2000 TSI Assay, IMMULITE 2000 TSI Calibration Verification Material
Siemens Healthcare Diagnostics, Inc., Mar 3, 2016