Recall Z-1834-2022 ยท DQX Wire, Guide, Catheter
Cook Incorporated
Fixed Core Wire Guide (Straight), Guidewire Reference Part Number/GPN: C-SF-18-30 G02261 C-SF-25-40 G01937 TSF-18-260 G00590 C-SF-25-50 G01938 TSF-35-180 G00652 C-SF-18-40 G02255 TSF-25-145 G00617 TSF-32-145 G00638 C-SF-15-50 G02274 TSF-18-145 G00587 TSF-18-50 G00593 TSF-35-145 G00650 TSF-21-145 G00602 THSF-35-180 G01237 TSF-35-80 G00664 C-SF-25-30 G02163 C-SF-21-30 G02351 THSF-35-80 G02408
| Date | Sep 28, 2022 (initiated Aug 16, 2022) |
| Status | Open, Classified |
| Reason | Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised. |
| Root cause | Under Investigation by firm |
| Action | On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235 |
| Quantity | 5,669 US; 5,286 OUS |
| Distribution | Domestic distribution nationwide. Foreign distribution worldwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K171764 | Fixed Core Wire Guides Cook Incorporated, Mar 9, 2018 |