Recall Z-1815-2022 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Olympus Corporation of the Americas

Olympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.

DateSep 26, 2022 (initiated Aug 12, 2022)
StatusOpen, Classified
ReasonAn out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
Root causeUnder Investigation by firm
ActionOlympus issued Urgent Medical Device Recall Letter on 8/12/22 to Surgical Department Risk Management. Letter states reason for recall, health risk and action to take: 1. Immediately locate any product with affected lot numbers listed in this communication. See Attachment 2 for pictures showing the location of lot numbers on the package. 2. Call your Olympus customer service representative at 1 800 848 9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product
Quantity44 units (5/box); 133 units (5/box) OUS
DistributionNationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212643POWERSEAL Curved Jaw Sealer and Divider, Double Action
Olympus Surgical Technologies America, Sep 27, 2021