ClearanceBrief
Recall Z-1777-2022 ยท KGN Wound Dressing With Animal-Derived Material(S)

Collagen Matrix Inc

Collagen Dental Wound Dressing - Bovine Dermis Tap, 25mm x 75mm Product was distributed under the following brand names: 1) Cytoplast RTMTAPE, 25mm x 75mm 2) GIBSON Healthcare, 25mm x 75mm 3) MaxxTape, 2.5cm x 7.5cm x 0.1cm 4) NEOTAPE, 25mm x 75mm 5) Karl Schumacher TAPE, 2.5cm x 7.5cm x 1mm

DateSep 23, 2022 (initiated Aug 3, 2022)
StatusOpen, Classified
ReasonThe packaging may not be sealed, which could compromise sterility.
Root causePackaging change control
ActionOn August 3, 2022, Collagen Matrix began informing the distributors (initial consignees) to place the lot in question on ship hold. On August 8, 2022, Collagen Matrix began informing distributors by email to return the concerned products. Consignees are asked to return any recalled product to the recalling firm and complete the Certificate of Acknowledgement within 15 days of receipt and return it via email.
Quantity1336 units
DistributionUS distribution to distributors in IL, IN, MI, NC, OH, and TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152600Collagen Dental Wound Dressings
Collagen Matrix, Inc., Jun 23, 2016