Recall Z-1645-2022 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

Artis Q floor- Model No. 10848280 Artis Q ceiling- Model No. 10848281 Artis Q biplane- Model No. 10848282 Artis Q zeego- Model No. 10848283 Artis Q.zen floor- Model No. 10848353 Artis Q.zen ceiling- Model No. 10848354 Artis Q.zen biplane- Model No. 10848355

DateAug 26, 2022 (initiated Aug 10, 2022)
StatusOpen, Classified
ReasonSiemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.
Root causeDevice Design
ActionOn August 10, 2022, the firm notified customers via "URGENT: MEDICAL DEVICE CORRECTION" Letter. Customers were advised that if the error occurs, the user should try to recover normal operation by a shutdown and restart of the system. Siemens will correct the error with a hardware modification via Update Instruction AX028/22/S and AX039/22/S. Siemens' service organization will contact customers shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. If you have any further questions, call 6
Quantity1,389 (590 US; 799 OUS) (Total)
DistributionWorldwide Distribution: US (nationwide) and OUS (Foreign) to countries of: United Arab Emirates; Armenia; Angola; Argentina; Austria; Australia; Bangladesh; Belgium; Bulgaria; Bahrain; Brazil; Canada; Switzerland; China; Colombia; Costa Rica; Czech Republic; Germany; Denmark; Dominican Republic; Alg