Recall Z-1770-2022 ยท FIP Subsystem, Water Purification

Fresenius Medical Care Holdings, Inc.

AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US

DateSep 19, 2022 (initiated Jun 21, 2022)
StatusOpen, Classified
ReasonInternal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
Root causeUnder Investigation by firm
ActionFresenius Medical Care Renal Therapies Group, LLC (FMCRTG) issued Urgent Medical Deice Correction letter (FA-2022-20-C) on 06/21/22 via First Class Mail. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) Water Systems Service Specialist will be contacting you to schedule a correction of your portable Water Treatment system. Only a Certified AquaC UNO H System Technician or FMCRTG Water Systems Service Specialist may perform this service. If choosing to correct via your own Certified AquaC UNO H System Technician, you w
Quantity6 units
DistributionUS Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143617AquaC UNO H
Vivonic GmbH, Oct 19, 2015