Recall Z-1766-2022 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Maquet Medical Systems USA

BEQ-HLS 5050 USA; HLS Set Advanced 5.0 Part No. 70106.9077. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.

DateJul 22, 2022
StatusOpen, Classified
ReasonThe firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).
Root causePackage design/selection
ActionOn July 15, 2022, the firm notified affected customers via Urgent Medical Device Removal Expansion letter. The letter informed customers of the inclusion of additional product codes and lots into the earlier recall initiated on November 6, 2020. Customers were instructed to examine their inventory and identify any affected product on hand. Affected product in use may be continued to be used according to normal practices. Unused and unexpired affected product should be immediately returned to Getinge for credit. Please contact Getinge Customer Service at (888) 9GETUSA / (888) 943-8872 (press
Quantity6,871 US; 18,670 worldwide
DistributionWorldwide distribution - US Nationwide and the countries Canada and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112360QUADROX-I MICROPORPOROUS MEMBRANE OXGENATOR SERIES AND QUDROX-ID DIFFUSION MEMBRANE OXYGENATORS SERIES
Maquet Cardiopulmonary, AG, Sep 15, 2011