Recall Z-1666-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Canon Medical System, USA, INC.

INFX-8000V. For radiographic and fluoroscopic studies and intervention.

DateJan 31, 2022
StatusOpen, Classified
ReasonThe fluoroscopic dose rate might exceed the conforming value during biplane fluoroscopy.
Root causeSoftware design
ActionCanon sent an Urgent Electronic Product Radiation Warning letter dated July 28, 2022 to customers. . Customer were informed that in the interim a temporary countermeasure will be performed until such time as the software becomes available. A representative of Canon Medical Systems USA will contact you to schedule a time to install the countermeasure and at a later date, the updated software to address the issue. For scheduling questions, please contact our InTouch team at intouchdispatch@us.medical.canon. For questions call 800-421-1968 or your Canon service representative at 800.521.1968.
Quantity16 systems
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170909Infinix INFX-8000V, V6.40
Toshiba Medical Systems Coroporation, Jun 16, 2017