Recall Z-0182-2023 ยท IZL System, X-Ray, Mobile

SEDECAL SA

Sedecal Soltus Battery Mobile X-Ray System REF Soltus 450 Battery Mobile X-Ray System REF Soltus 500

DateNov 4, 2022 (initiated Aug 12, 2022)
StatusOpen, Classified
ReasonDue to un-commanded movement of mobile x-ray unit.
Root causeSoftware design
ActionSedecal SA provided Canon Medical Systems USA (CMSU) the sole distributor of the Soltus system in the U.S. an "URGENT: ADVISORY NOTICE" to communicated to CMSU's customers. On 07/28/2022, CMSU sent the "URGENT: ADVISORY NOTICE" Letter via USPS Certified to inform customer of an issue with erratic movements related with Soltus 500 Mobile X-Ray units. According to reports, the users were trying to turn the systems to the left or to the right when it suddenly started moving backwards at a high speed. Customer are instructed to: 1. Do not continue using the unit when in error condition (magenta
Quantity155 systems
DistributionU.S.: KY, WI, FL, DE, WV, CA, MO, NY, NC, AL, AR, MS, GA, IL, IA, TN, KS, KY, IN, HI, OH, WY, PA, NV, RI, UT, AZ, OK, MN, ME, MT, WA, VA, SD, MI, LA, and TX O.U.S.: N/A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192936Soltus 500
Sedecal., Sa., Nov 15, 2019