Recall Z-1795-2022 ยท FYD Apparatus, Exhaust, Surgical
Mckesson Medical-Surgical Inc. Corporate Office
Cooper Surgical PREFILTER, SMOKE EVAC (10/BX)CPRSRG Model Number: 52560.
| Date | Sep 26, 2022 (initiated May 25, 2022) |
| Status | Open, Classified |
| Reason | Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. |
| Root cause | Environmental control |
| Action | Mckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 sedgwick@sedgwick.com. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to pr |
| Quantity | 9 boxes |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K983298 | BUFFALO FILTER VIROSAFE ERBIUM:YAG FILTER - #VSEYAG1-4 BUFFALO FILTER VIROSAFE ERBIUM:YAG PREFILTER - #BFEYAG04PF Buffalo Filter Co., Inc., Aug 16, 1999 |