Recall Z-1563-2022 ยท KPR System, X-Ray, Stationary

Philips North America Llc

DigitalDiagnost C50 1.1

DateSep 1, 2022 (initiated Feb 25, 2022)
StatusOpen, Classified
ReasonFollowing a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Added 1/3/2023), Expansion of the root cau
Root causeSoftware Design Change
ActionPhilips notified customers of the issue via an Electronic Product Radiation Defect notification letter dated July 12, 2022. Philips will schedule an appointment with customers to install the software update at no cost to customers. For questions contact Philips Customer Care Solutions Center 1-800-722-9377.
Quantity11 systems (updated 1/3/2023)
DistributionNationwide Distribution including Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163410DigitalDiagnost C50
Philips Healthcare (Suzhou) Co., Ltd., Jan 4, 2017
K201725DigitalDiagnost C50
Philips Healthcare (Suzhou) Co., Lit., Jul 10, 2020