Recall Z-1855-2022 ยท BSY Catheters, Suction, Tracheobronchial

MEDLINE INDUSTRIES, LP - Northfield

MEDLINE CLEARPRO DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. ET LENGTH DOUBLE SWIVEL, 12 Fr, Item Number/REF: DYNCPDS12; b. ET LENGTH DOUBLE SWIVEL, 14 Fr, Item Number/REF: DYNCPDS14; c. ET LENGTH DOUBLE SWIVEL, 16 Fr, Item Number/REF: DYNCPDS16

DateSep 29, 2022 (initiated Aug 3, 2022)
StatusOpen, Classified
ReasonPotential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
Root causeUnder Investigation by firm
ActionOn about 08/03/2022, Medline notified consignees via first class mail letter of the recall. The letter instructed consignees to immediately review inventory and destroy all affected product and complete the response form indicated at the link provided in the letter. Distributors were instructed to notify customers if further distributed, instruct those customers destroy any affected product, and include customer quantities on response form.
Quantity65,320 units
DistributionNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Pa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.