Recall Z-1742-2022 ยท IYE Accelerator, Linear, Medical

Zap Surgical Systems

ZAP-X Radiosurgery System

DateJul 5, 2022
StatusOpen, Classified
ReasonSoftware issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.
Root causeSoftware design
ActionOn 7/05/22 correction notices were distributed to customers who were told that revised software to address the recall issue would be supplied. Until the firm supplies the revised software follow the following steps to ensure accurate treatment delivery: 1) After using auto-align ensure initial patient setup (prior to treating the first isocenter), has rotational deviations of less than plus-minus 1.5 degrees in each dimension. 2) Prior to commencement of each isocenter beam delivery, verify that rotational deviations remain below plus-minus 1.5 degrees. - Additionally, visually verify patien
Quantity8
DistributionWorldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211663Zap-X Radiosurgical System
Zap Surgical Systems, Inc., Jul 29, 2021