Recall Z-1524-2022 ยท QJR Reagents, 2019-Novel Coronavirus Nucleic Acid

Luminex Corporation

ARIES SARS-CoV-2 Assay

DateJul 14, 2022
StatusOpen, Classified
ReasonSARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not detected, a false negative result would be generated.
Root causeNonconforming Material/Component
ActionOn 7/14/22 recall notices were emailed to consignees who were asked to review any results from the affected lot to confirm the validity of the results. In addition, all remaining cassettes from this lot should be discarded. Customers with questions are encouraged to contact the firm at: 1-877-785-2323 (U.S. and Canada), +1-512-381-4397 (Outside U.S. and Canada), support@luminexcorp.com
Quantity257
DistributionU.S. Nationwide distribution in the states of NE, TX, MS, TN, KY, ME, CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.