ClearanceBrief
Recall Z-1655-2022 ยท MNS Ventilator, Continuous, Non-Life-Supporting

Philips Respironics, Inc.

Philips BiPAP A40, Rx Only. To provide noninvasive ventilation support for patients with obstructive sleep apnea syndrome (OSA) and respiratory insufficiency.

DateSep 9, 2022 (initiated Aug 26, 2022)
StatusOpen, Classified
ReasonMotor assemblies manufactured with nonconforming plastic that contributes to off-gassing of VOCs.
Root causeProcess control
ActionThe firm distributed an "URGENT Medical Device Recall" letter to customers dated 8/26/2022 via mail. The notification informs customers that certain BiPAP V30 Auto and OmniLab Advanced+ units distributed in the US have the potential to off-gas harmful chemicals and structural failure of the device motor due to nonconforming material used in the assembly. Patients are to stop using the recalled devices and consult with their physician to determine the most appropriate options for continued treatment. Customers are asked to locate all affected devices and immediately stop use and remove them f
Quantity1,091 units ***updated Jan 2024*** additional 323 units
DistributionDomestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, TN, TX, VA, VT, WA, WI, WV, & WY. Foreign: Australia, Austria, Bangladesh, Brazil, Canada, China, Denmark, Egypt, France, Germany, Hong Kong, Hungary, In
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090248BIPAP AUTOSV
Respironics Inc., Sleep & Home Respiratory Group, Mar 4, 2009