ClearanceBrief
Recall Z-1660-2022 ยท MMI Immunoassay Method, Troponin Subunit

Abbott Laboratories

ARCHITECT STAT Troponin-I Reagent Kit, REF 2K41-27

DateJul 26, 2022
StatusOpen, Classified
ReasonOne lot of reagent was manufactured with insufficient quantities of the heterophilic blocking agents in the conjugate component.
Root causeUnder Investigation by firm
ActionOn July 26, 2022, Abbott issued a Product Recall letter for this issue sent the letter via FedEx priority overnight express. The letter provided the reason for the recall and hazard involved and requested the consignee take the following actions: "" IMMEDIATELY discontinue use of ARCHITECT STAT Troponin-I, List Number 2K41-27 lot number 63718UN22 and destroy all inventory of lot 63718UN22 according to your laboratory procedures. " Please contact Customer Support to order replacement material. " Please review this letter with your Medical Director or Laboratory Management and follow your labor
Quantity2451 kits of 100 tests
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041192ARCHITECT STAT TROPONIN-I IMMUNOASSAY
Fisher Diagnostics, Aug 12, 2004