Recall Z-1651-2022 ยท OWD Somatic Gene Mutation Detection System

Qiagen Sciences LLC

therascreen FGFR RGQ RT-PCR Kit (CE IVD, not sold in the US) REF 874711

DateJul 20, 2022
StatusOpen, Classified
ReasonIncreased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive result could lead to the risk of the patient being exposed to an inappropriate or suboptimal anti-cancer treatment and/or unnecessarily experiencing side effects
Root causeUnder Investigation by firm
ActionQiagen issued July 20,2022 letter ( REF 874721) via email (A) US customers: notify customers by Urgent Medical Device Correction Letter; (B) CE customers: notify customers by Urgent Field Safety Notice (C) RUO (not IVD) customers: notify customers by Important Notice. Letter states reason for recall, health risk and action to take: For the continued use of REF 874721 therascreen FGFR RGQ RT-PCR Kit LOT 172017274 , please note the following criteria: Samples with the Overall sample result No Alteration Detected should be regarded as correct and no further actions are needed. Samples with t
Quantity29 kits
DistributionIN, NC, MN, TX Foreign: Belgium, China, Germany, Jordan, Italy, Netherlands, Malaysia, Spain, Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.