Recall Z-1571-2022 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vanta implantable neurostimulators for pain therapy.

DateJul 26, 2022
StatusOpen, Classified
ReasonPotential for Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 to display an Unexpected Device Error Code 1502 Message, and the user will be unable to perform programming of the Vanta Implantable Neurostimulator (INS) Model 977006.
Root causeUnder Investigation by firm
ActionOn about 07/26/2022, Medtronic notified consignees via mail courier service with a letter titled "URGENT: MEDICAL DEVICE CORRECTION." The letter instructed consignees to perform an initial interrogation of the Vanta INS and tap start usage, tap implant device workflow, click start, and on the next screen exit the session, and to initiate a second interrogation and if there is no "Unexpected Device Error Code 1502" code the INS maybe used. If the INS cannot be interrogated, contact Medtronic Technical Services at 1-800-707-0933. Additionally, consignees were instructed to complete and
Quantity2,920 downloaded software applications
DistributionWorldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. The countrie
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.