Recall Z-1574-2022 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vanta implantable neurostimulators for pain therapy.

DateAug 25, 2022 (initiated Jul 26, 2022)
StatusOpen, Classified
ReasonPotential communication issue in which the Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 may be unable to connect with a Vanta Implantable Neurostimulator (INS) Model 977006 when a patient is implanted with more than one neurostimulator.
Root causeUnder Investigation by firm
ActionOn about 07/26/2022, Medtronic notified consignees via letter titled "URGENT: MEDICAL DEVICE CORRECTION" sent through courier service. The notification instructed consignees to hold off implanting more than one Vanta INS in a patient until software update is available and while not recommended, if implanting a patient with two (2) Vanta INSs, ensure the INSs are implanted on opposite sides of the body as noted in the Information for Prescribers labeling, and ensure each is configured with the CP App before implanting in a patient. Consignees were asked to complete and return the customer con
Quantity4386 downloaded software applications
DistributionWorldwide distribution - US distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico. The countries of Aus
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.