ClearanceBrief
Recall Z-1676-2022 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Ortho-Clinical Diagnostics, Inc.

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

DateJul 19, 2022
StatusOpen, Classified
ReasonPerforming a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.
Root causeUnder Investigation by firm
ActionOrtho CLinical issued Urgent Medical Device Correction letter on 19 July 2022 to consignees who received a VITROS XT3400 or a VITROS XT7600 System notified via (CL2020-176) of the issue and instructed to: " If your VITROS XT 3400 and/or VITROS XT 7600 Systems require a system shutdown, routine Quality Control processing for MicroSlide assays must be performed per your laboratory procedure, prior to processing patient samples, to determine if the anomaly has occurred. " If QC results are outside your established ranges, Ortho recommends performing a second system shutdown and re-processing QC
Quantity1148 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182063VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS Chemistry Products TBIL Slides, VIT
Ortho-Clinical Diagnostics, Inc., Oct 30, 2018