Recall Z-1675-2022 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Ortho-Clinical Diagnostics, Inc.
VITROS XT 3400 Chemistry System with Version 3.7.2 and below Product Code: 6844458
| Date | Jul 19, 2022 |
| Status | Open, Classified |
| Reason | Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System. |
| Root cause | Under Investigation by firm |
| Action | Ortho CLinical issued Urgent Medical Device Correction letter on 19 July 2022 to consignees who received a VITROS XT3400 or a VITROS XT7600 System notified via (CL2020-176) of the issue and instructed to: " If your VITROS XT 3400 and/or VITROS XT 7600 Systems require a system shutdown, routine Quality Control processing for MicroSlide assays must be performed per your laboratory procedure, prior to processing patient samples, to determine if the anomaly has occurred. " If QC results are outside your established ranges, Ortho recommends performing a second system shutdown and re-processing QC |
| Quantity | 506 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K182063 | VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS Chemistry Products TBIL Slides, VIT Ortho-Clinical Diagnostics, Inc., Oct 30, 2018 |