Recall Z-1565-2022 ยท DZE Implant, Endosseous, Root-Form

Nobel Biocare

Nobel Biocare N1 TiUltra TCC NP 3.5x13mm

DateAug 23, 2022 (initiated Jul 25, 2022)
StatusOpen, Classified
ReasonDue to the presence of a burr which may cause an aspiration risk to patient during surgical procedure.
Root causeProcess change control
ActionOn 07/25/2022, Nobel Biocare AB sent an "URGENT MEDICAL DEVICE RECALL" Letter via UPS to customers informing them that there may be a burr present on the co-packed OsseoShaper 1 around the waist of the device. The hazard is primarily identified as a risk of the patient swallowing or inhaling the burr if it comes loose from the OsseoShaper 1 during the clinical procedure. Customer are instructed to: 1. Inspect their stock and quarantine affected devices. 2. Complete attached Customer Acknowledgment Form, even if they do not have any affected stock, and return it to Nobel Biocare, via email to
Quantity347 devices
DistributionU.S Nationwide.: AZ, CA, CT, FL, HI, NC, NY, OR, PA, TX, UT, and WI O.U.S.: Austria, Belgium, Canada, Germany, Spain, France, United Kingdom, Hungary, Iceland, Italy, Netherlands, Norway, Portugal, and Sweden
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211109N1 TiUltra TCC Implant System
Nobel Biocare Services AG, Dec 21, 2021