ClearanceBrief
Recall Z-1649-2022 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

New Standard Device Inc

Metalogix Self-Drill Half-Pin, a component of the Revolution External Fixation System. Orthopedic. Metalogix Self-Drill Half-Pin 5x30x215mm REF 620030 Metalogix Self-Drill Half-Pin 5x40x215mm REF 620040 Metalogix Self-Drill Half-Pin 5x60x215mm REF 620060 Metalogix Self-Drill Half-Pin 5x80x215mm REF 620080

DateDec 28, 2020
StatusOpen, Classified
ReasonComplaints were received for Self-drill Half-Pin tip breakages occurring during surgery.
Root causeDevice Design
ActionOn 12/28/2020, Metalogix send a "Field Notice: Self-Drill Half-Pin" via email. The Notice only informed customer that the Citieffe Half-Pins will temporarily be used to replace the current 5mm Self-Drill Half-Pins. A contact number for Mr. Elton Ely, QA Manager was provided in the notice 833-659-2019 ext. 6
Quantity410 total devices
DistributionU.S.: AZ, GA, IL MT, NC, NJ, NV, OH, PA, TX, and VA O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181630Revolution External Plating System
New Standard Device Dba Metalogix, May 29, 2019