Recall Z-1515-2022 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Operations, Inc.

cobas e801 Immunoassay Analyzer

DateJul 14, 2022
StatusOpen, Classified
ReasonSoftware issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.
Root causeSoftware design
ActionThe firm distributed an Urgent Medical Device Correction notification via mail by UPS on 07/14/2022. The notice indicates that cobas e801 (Catalog No. 08454345001) and cobas e801 module (Catalog No. 04682913001) are being recalled due to a software issue affecting all analyzers, which may result in impacted signals and sample results of Pre-Wash tests when using Quick Start Mode for the following assays: Anti-HAV 2, Anti-HBc IgM, IGF-1, Myoglobin, Rubella IgG, and Toxo IgG. Customers with affected units are provided workaround instructions to deactivate Quick Start Mode or additional software
Quantity965 units
DistributionUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, DC & PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.