Recall Z-1677-2022 ยท DYB Introducer, Catheter

Biosense Webster, Inc.

CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath, REF: D138501. Used to introduce various cardiovascular catheters into the heart.

DateJun 29, 2022
StatusOpen, Classified
ReasonOutside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised sterile barrier could be a source for infection.
Root causeMaterial/Component Contamination
ActionOn 6/29/22, re call notices started being emailed and mailed to customers who were asked to do the following: 1) The notice should be forwarded to anyone in the facility that needs to be aware. 2) Identify affected devices, unbox devices, and destroy the devices. 3) Complete and return the Business Reply Form and send back the label on empty boxes to receive credit. 4) If any of the affected products have been forwarded to another facility, contact that facility, and provide a copy of this letter to the relevant personnel. 5) Maintain a copy of this communication where the product identified i
Quantity98
DistributionUS Nationwide distribution in the states of AL, GA, AZ, TX, MA, VA, NY, ME, FL, IA, NM, NH, TN, WI, OH, SC, CA, IN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170997CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath
Biosense Webster, Inc., Jul 25, 2017