Recall Z-1732-2022 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Exactech, Inc.

GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS, REF 130-32-51 b. NOVATION GXL LINER NEUTRAL, 36MM ID, GROUP 2 CUPS, REF 130-36-52 c. NOVATION GXL LINER NEUTRAL, 40MM ID, GROUP 3 CUPS, REF 130-40-53 d. NOVATION GXL LINER NEUTRAL, 40MM ID, GROUP 4 CUPS, REF 130-40-54 e. NOVATION GXL LINER NEUTRAL, 40MM ID, GROUP

DateAug 11, 2022
StatusOpen, Classified
ReasonSpecific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
Root causeProcess control
ActionThe firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providi
Quantity17929 units
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON PORTUGAL,PUERTO RICO,SPAIN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121392NOVATION CROWN CUP LINERS; NEUTRAL, LIPPED, +5MM LATERALIZED, +5/10 DEGREE
Exactech, Inc., Jan 16, 2013